Vesscide Cutting Balloon Dilatation Catheter
The Vesscide Cutting Balloon Dilatation Catheter has the following performance characteristics:
(1). Semi-compliant balloon platform is adopted in the design of the product. Semi-compliant balloon has good flexibility, this will bring good retraction performance and lesion-crossing ability, and control the degree to which the blade is embedded in the calcified plaque during inflation, providing a strong guarantee for surgical safety.
(2). The cutting blades are made of stainless steel, which has high hardness, high strength and high wear resistance. The blades adopt precise geometric shapes and sharp edges to ensure an effective lesion cutting performance. Besides, blade is extremely thin with a safe height of 0.15mm it only cuts the inner membrane of the blood vessel wall and carves a mark on the inner wall of the blood vessel, and does not cut the entire blood vessel wall, avoiding damage to the normal blood vessel wall. The blades are arranged in parallel on the surface of the balloon and are firmly fixed to the balloon through a specific process. When the balloon is inflated, these blades will generate evenly distributed pressure on the lesion to achieve precise cutting performance.
(3). Multiple specifications of the device (balloon diameter * balloon length * number of blade subassembly) are available for different lesion conditions.
(4). The adoption of hydrophilic coating brings good pushability and lesion-crossing ability.
(5). Platinum marker of the device makes it clearly visible under X-ray operation.
1. Adapted guiding catheter: Minimum guiding catheter 1.8mm/6F
2. Adapted guiding wire:0.36mm/0.014inch
3. Effective length of the catheter:1350mm
4. Marker location:Both ends of the balloon
The Vesscide Cutting Balloon Dilatation Catheter is intended for use in the treatment of coronary artery stenosis. The target lesion should possess the following characteristics:
(1) Localized lesions with lesion length ≤ 20mm.
(2) Reference vessel diameter (RVD) of 2.00 mm to 4.00 mm
(3) Readily accessible to the device
(4) Light to moderate tortuosity of proximal vessel segment
(5) Non-angulated lesion segment (< 45° )
(6) Smooth angiographic contour
(7) Absence of angiographically visible thrombus.