VessTec™ Disposable Pressure Micro-catheter & KF12 Fractional Flow Reserve Tester
Diagram and Technical Parameters


Technical parameters
VessTec™ Disposable Pressure Micro-catheter
Model & specification: LPPMC-RX26
Pressure measuring range: -30~+300 mmHg
Excitation voltage: DC 2.5 V
Excitation impedance: >200 Ω
Signal output impedance: < 4,000 Ω
Zero drift: ±7 mmHg
Accuracy: less than ±1 mmHg plus ±1% of the reading
(for pressure range: -30 ~ +50 mmHg), ±3% of the
reading (for pressure range: 50 ~ 300 mmHg)
Nominal sensitivity: 3-10 uV/V/mmHg
Luminous sensitivity:<5 mmHg
Operating temperature: 15~40°C
KF12 Fractional Flow Reserve Tester
Display screen: 12.1", LED
Unit dimensions: 313X178X334 (mm)
Net weight: 5.2 KG (including 1 battery)
Continuous working hours: ≥3H (fully charged)
Pa measuring range: -30 mmHg ~ 300 mmHg
Pd measuring range: -30 mmHg~300 mmHg
Average cycle setting of Pa and Pd: 1b, 2b, 3b, 8b and 12b
FFR waveform scanning speed: 5 mm/s, 10 mm/s,
25 mm/s and 50 mm/s
The setting of recording time: 10 min, 20 min and 30 min
Product Features

Clinical study
Evaluation Criteria : evaluating the diagnostic accuracy of FFRPMC with FFRPW as the golden standard
The diagnostic accuracy of FFRpmc is up to 97.02%, and the lower
confidence limit is greater than the preset target value.

Key Points

Thank You for Your Attention on Lepu Medical!
Email us with any questions or inquiries or use our contact data. We would be happy to answer your questions.